Reversed-Phase Liquid Chromatographic Method for Simultaneous Determination of Artesunate and Mefloquine in Pharmaceutical Preparations
T.M. Kalyankar1* and R.B.Kakde2.
1School of Pharmacy, S.R.T.M. University, Nanded, Maharashtra, India.
2Departments of Pharmaceutical Sciences, R.T.M. Nagpur University, Nagpur, Maharashtra, India.
*Corresponding Author E-mail: kalyankarr@rediffmail.com
ABSTRACT:
A simple, rapid, precise and accurate reversed phase high performance liquid chromatographic method has been developed for the simultaneous determination of Artesunate in combination with Mefloquine. This method uses a Hypersil ODS C18 (250mm×4.6mm×5µ particle Size) analytical column, a mobile phase of Methanol, Acetonitrile and 0.1% Acetic acid in the ratio 70:10:20 (v/v), and pH 4.5 adjusted with orthophosphoric acid. The instrumental settings are a flow rate of 1 ml/min and PDA detector wavelength at 210-360 nm. The retention times for Artesunate and Mefloquine are 6.57 min and 2.19min, respectively. The method is validated and shown to be linear. The linearity range for Artesunate and Mefloquine are 250-1500 and 500-3000 µg/ml respectively. The Percentage recovery for Artesunate and Mefloquine are 99.60-99.35 and 99.49-99.78 respectively. The correlation coefficients of Artesuante and Mefloquine are 0.999, and 0.999, respectively. The relative standard deviation for six replicates is always less than 2%. The Statistical analysis proves that the method is suitable for analysis of Artesunate and Mefloquine as a bulk drug and in pharmaceutical formulation without any interference from the excipients
KEYWORDS: Artesunate, Mefloquine, RP-HPLC and validation
INTRODUCTION:
Artesunate (Fig.-1A) chemically is (3R,5aS,6R,8aS,9R,10S,12R,12aR)–decahydro-3,6,9-trimethyl-3,12-epoxy-12H-pyrano[4,3-j]-1,2-benzodioxepin-10-o1,hydrogensuccinate . Artesunate and its active metabolite dihydroartemisinin are potent blood schizonticides, active against the ring stage of the parasite1.
Mefloquine (Fig.-1B) chemically is a 2-piperidinyl-2,8-bis(trifluoromethyl)-4-quinolinemethanol hydrochloride is a recently developed antimalarial agent. It is effective as a single oral dose against multidrug-resistant malaria in man2. It also provides suppressive prophylaxis against Plasmodium vivax and P. falciparum infections in human volunteers3.
Literature survey shows that very few methods are reported for the individual estimation of Artesunate and Mefloquine. The individual methods reported for Artesunate4-9 and Mefloquine10-13 are only for the estimation of these drugs in tablet and plasma. However there is no single method reported for the simultaneous determination of Artesunate and Mefloquine.
In the present research work, attempts have been made to develop a simple, sensitive method for the simultaneous determination of Artesunate and Mefloquine.
A
B
Figure 1: Structures of Antimaleria Drugs: A-Aretsunate, B-Mefloquine
EXPERIMENTAL:
Instrumentation
The LC system was from Perkin Elmer quaternary pump Series 200 and was comprised of auto sampler injector; and an Intelligence PDA detector connected to the Total Chrome Navigator version 6.3. For controlling the instrumentation as well as processing the data generated was used.
Material and reagents
Authentic samples of Artesunate (ASN) and Mefloquine (MEF) both were obtained from Macleod Pharmaceuticals Ltd (Mumbai, India) used as such without further purification. Brand of tablets FALCIGO Plus (Zydus Cadila, India), labeled to contain 100 mg of ASN 200 mg Mefloquine were procured from the local market. Acetonitrile (HPLC grade) were obtained from Rankem Pvt. Ltd. (Delhi, India). Acetic acid (HPLC grade), Methanol (HPLC grade), orthophosphoric acid (AR grade) and double distilled water were used in analysis.
Chromatographic conditions
The optimum composition of the mobile phase containing Acetonitrile: methanol and 0.1% Acetic acid (10:70:20 v/v) of pH 4.5 adjusted with orthophosphoric acid, was selected because it was found to ideally resolve the peaks of ASN and MEF. The flow rate was set to 1.0 ml/min and UV detection was carried out at 210 to 360 nm, separation is carried out by c18 column (5µm; 250cm x 4.6mm) temperature was maintained at ambient temperature. The data was acquired, stored and analysed with total chrome software.
Mobile phase
The mobile phase consisted of Acetonitrile: methanol and 0.1% Acetic acid (10:70:20 v/v) of pH 4.5 adjusted with orthophosphoric acid. The mobile phase was degassed with ultra sonication.
Standard preparation
Artesunate
An accurately weighed 500 mg of Artesunate was transferred in a 50 mL volumetric flask, and dissolved with methanol. The solution was sonicated for 20 min and then volume was made upto the mark.
Mefloquine
An accurately weighed 500 mg of mefloquine was transferred in a 50 mL volumetric flask, and dissolved with methanol. The solution was sonicated for 20 min and then volume was made upto the mark.
Calibration curve solutions
From the mentioned stock solutions of ASN and MEF calibration curve solutions containing 250-1500µg/mL of ASN, 500-3000µg/mL of MEF in each calibration level were prepared.
Preparation of sample solutions
Twenty tablets were weighed and finely powdered. A quantity equivalent to one tablet containing 100 mg of ASN and 200 mg of MEF was transferred in a 50 mL volumetric flask and volume was made with methanol. The contents were sonicated with mobile phase for 20 min, and filtered through 0.45μm membrane filter.
RESULTS AND DISCUSSION:
Optimization of chromatographic conditions
The chromatographic method was optimized by changing the various variable parameters of the mobile phase. Different experiments were performed to achieve the adequate retentions and resolution for the peaks of Artesunate and Mefloquine. To set the adequate retentions and resolution, the effects of the mobile phase components, changes in ionic strength, and pH of the mobile phase were studied. From the pH effects study, it was observed that the resolution increased with decreasing the pH of the mobile phase, and at pH 3.5, the peaks of Artesunate and Mefloquine merge with each other. From the described studies, it was observed acitonitrile and methanol and 0.1% acetic acid (pH 4.5 adjusted with OPA) in the ratio of 10:70:20, v/v, gave adequate retentions and resolution, and the chromatographic run was less than 15 min.
Validation of the method
Specificity
The specificity of the method was checked by a peak purity test of the sample preparation done by PDA detector. The peak purity for ASN and MEF was found to be 999. The result of the peak purity analysis shows that the peaks of the analytes were pure and also the formation excipients were not interfering with the analyte peaks.
Calibration and linearity
The standard solutions containing 250 - 1500 µg/ml of ASN and 500-3000 µg/ml of MEF in each linearity level were prepared. Linearity solutions were injected in triplicate. In the simultaneous determination, the calibration graphs were found to be linear for both the analytes in the mentioned concentrations. The coefficient of correlation was found to be 0.999 and 0.999 for ASN and MEF, respectively.
Precision (repeatability)
The precision of the method was studied by determining the concentrations of each ingredient in the tablets six times. In the precision study, relative standard deviation of the ASN and MEF were found to be 0.01791and 0.2087respectively. The results of precision study indicate that the method is reproducible.
Accuracy (recovery test)
The accuracy of the method was studied by recovery experiments. The recovery experiments were performed by adding known amounts of the pure drug. The recovery was done at three levels: 80%, 100%, and 120% of the label claim. Three samples were prepared for each recovery level.
Table I. Results of the Recovery Tests for the Drugs (n = 3)
|
Level of addition (%) |
Amount added (mg ) |
Recovery (%)* |
Average Recovery† |
|||
|
ASN |
MEF |
ASN |
MEF |
ASN |
MEF |
|
|
80 |
80 |
160 |
99.60(0.555) |
99.49(0.3667) |
99.73 (0.6963) |
99.86 (0.5349) |
|
100 |
100 |
200 |
100.26(0.6222) |
100.33(0.778) |
||
|
120 |
120 |
240 |
99.35(0.9118) |
99.78(0.4602) |
||
* RSD shown in parenthesis.
† Average recovery = average of three levels, nine determinations
Table II. Assay Results of Active Ingredients in Tablets
|
Set |
Ingredients |
Label value (mg) |
Found (mg) † |
% Label claim |
|
Precision |
ASN |
100 |
100.2 |
100.2(0.798) |
|
MEF |
200 |
200.21 |
100.10(0.096) |
|
|
Intermediate precision |
ASN |
100 |
99.89 |
99.86(0.577) |
|
MEF |
200 |
200.01 |
100.00(0.255) |
† Average of six analyses
The recovery values for ASN and MEF ranged from 99.60-99.35 and 99.49-99.78, respectively (Table I). The average recoveries of three levels for ASN and MEF were found to be 99.73 and 99.86 respectively.
Intermediate precision
Intermediate precision of the method was done by analyzing the sample six times on different days, by different chemists, using different analytical column of the make, and different HPLC systems. The percentage assay was calculated using the calibration curve. The assay results are shown in Table II.
Determination of the limits of detection and Quantitation
For determining the limits of detection (LOD) and Quantitation (LOQ), the method based on the residual standard deviation (SD) of a regression line and slope was adopted. To determine the LOD and LOQ, a specific calibration curve was studied using samples containing the analytes in the range of the detection and quantitation limits. The LOD for ASN and MEF were 0.00923 and 0.00190 µg/mL, and the LOQ were 0.02795 and 0.0057721 µg/ml respectively.
System suitability
For system suitability studies, six replicate injections of mixed standard solutions were injected, and the parameters like RSD of peak area ratio, column efficiency, resolution, and tailing factor of the peaks were calculated. Results are shown in Table III.
Table III. System Suitability Parameters
|
Parameters |
ASN |
MEF |
|
Retention time (min) |
6.57 |
2.19 |
|
Capacity factor |
1.49 |
1.26 |
|
Tailing Factor |
1.16 |
1.24 |
|
Theoretical Plates |
10466 |
5678 |
|
Resolution |
11.86 |
|
Robustness
To evaluate robustness of the developed method, few parameters were deliberately varied. These parameters included variation of flow rate, percentage of Acetonitrile in the mobile phase, pH of mobile phase and temperature. Each factor selected was changed at three levels (-1, 0, +1). One factor was changed at one time to estimate the effect. The results are shown in Table IV.
Determination of active ingredients in tablets
The contents of two drugs in tablets were determined by the proposed method using a calibration curve. The determinations were done in two sets, one for precision and the second for intermediate precision, and six samples were prepared for each set. The results are shown in table III. The chromatogram of the tablet sample is shown in figure2.
Figure 2. A typical chromatogram of Artesunate and Mefloquine.
CONCLUSIONS:
This method can be used for the simultaneous determination of Artesunate and Mefloquine in the pharmaceutical dosage form. The method is validated and shown to be accurate and precise. It can be used in the quality control departments for the analysis of Artesunate in combination with Mefloquine.
Table IV. Summary of Robustness Study
|
Chromatographic changes |
|||||||||||||
|
Flow Rate (ml/min) |
Level |
RT |
Tailing factor |
% Content found |
|||||||||
|
ASN |
MEF |
ASN |
MEF |
ASN |
MEF |
||||||||
|
0.7 |
-2 |
6.58 |
2.22 |
1.20 |
1.21 |
99.55 |
98.40 |
||||||
|
0.8 |
0 |
6.57 |
2.19 |
1.16 |
1.24 |
99.89 |
100.02 |
||||||
|
0.9 |
+2 |
6.01 |
2.03 |
1.19 |
1.26 |
99.54 |
99.96 |
||||||
|
Mean* ± S.D |
6.55± 0.0208 |
2.17± 0.023 |
|
||||||||||
|
pH of the mobile phase |
Level |
RT |
Tailing factor |
% Content found |
|||||||||
|
ASN |
MEF |
ASN |
MEF |
ASN |
MEF |
||||||||
|
4.4 |
-2 |
6.52 |
2.16 |
1.19 |
1.22 |
98.90 |
101.88 |
||||||
|
4.6 |
0 |
6.57 |
2.19 |
1.16 |
1.24 |
99.89 |
100.02 |
||||||
|
4.8 |
+2 |
6.56 |
2.21 |
1.12 |
1.19 |
101.65 |
101.56 |
||||||
|
Mean* ± S.D |
6.57±0.024 |
2.19±0.021 |
|
||||||||||
|
Temperature |
Level |
RT |
Tailing factor |
%Content found |
|||||||||
|
ASN |
MEF |
ASN |
MEF |
ASN |
MEF |
||||||||
|
29 |
-1 |
6.53 |
2.15 |
1.13 |
1.28 |
100.03 |
100.26 |
||||||
|
30 |
0 |
6.57 |
2.19 |
1.16 |
1.24 |
99.89 |
100.02 |
||||||
|
31 |
+1 |
6.58 |
2.20 |
1.10 |
1.19 |
99.90 |
99.05 |
||||||
|
Mean* ± S.D |
6.57±0.023 |
2.17±0.012 |
|
||||||||||
|
Mobile phase concentration |
Level |
RT |
Tailing factor |
%Content found |
|||||||||
|
ASN |
MEF |
ASN |
MEF |
ASN |
MEF |
||||||||
|
68:08:24 |
-1 |
6.56 |
2.18 |
1.11 |
1.26 |
100.03 |
100.26 |
||||||
|
70:10:20 |
0 |
6.57 |
2.19 |
1.16 |
1.24 |
99.89 |
100.02 |
||||||
|
72:12:16 |
+1 |
6.53 |
2.15 |
1.15 |
1.21 |
99.90 |
99.05 |
||||||
|
Mean* ± S.D |
6.55±0.028 |
2.17±0.0145 |
|
||||||||||
* Mean of three levels (n = 3)
ACKNOWLEDGMENTS:
The authors are grateful to the School of Pharmacy S.R.T.M. University Nanded, Maharashtra State, India for providing the facilities for this research work.
REFERENCES:
1. Ranhar SS, Gandhi SV, Kadukar SS and Ranjane, A Validation HPLC method for determination of Artesunate in bulk and tablet formulation, Journal of Analytical Chemistry, 65(5); 2010: 507-510.
2. World Health Organization, Antimalarial Drug Combination Therapy. Report of a WHO Technical Consultation, Geneva, WHO (2001)
3. Kevin T. Batty, Timothy ME, Devis, Le Thi Anh Thu, Tran Quang Binh, and Trinh Kim Anh, Kenneth F, Selective HPLC determination of Artesunate and β- dihydroartemisinin in patients with falciparum malaria, Journal of Chromatography B, 677; 1996: 345-350.
4. Na-Bangchang K, Congpuong K, Hung LN, Mounto P, Karbwang J, Simple HPLC method with electrochemical detection for the simultaneous determination of Artesunate and dihydroartemisinin in biological fluids, Journal of Chromatography B 708; 1998: 201-207.
5. Choon-Sheen Lai, Nair NK, Mansor SM, Olliaro PL and Naratnam V, An analytical method with a single extraction procedure and two separate HPLC systems for the determination of Artesunate and dihydroartemisinin and Mefloquine in human plasma for application in clinical pharmacological studies of the drug combination, Journal of chromatography B 857; 2007: 308-314.
6. Bonnie A. Avery, Krishna K. Venkatesh and Mitchell A. Avery, Rapid determination of artemisinin and related analogues using HPLC and evaporative light scattering detector Journal of Chromatography B, 730; 1999: 71-80.
7. Rao Aswin B and Murthy RSR, A rapid spectrophotometric method for the determination of Mefloquine hydrocxhloride, Journal of Pharmaceutical and Biomedical Analysis, 27; 2002: 959-965.
8. Wallen Leif, Ericsson Orjan, Wikstrom Inger and Hellgren, HPLC method for the enantioselective analysis of Mefloquine in plasma and urine, Journal of chromatography B, 655; 1994: 153-157.
9. Bergqvist Vngve, Eckerbom Solveig, Larsson Helena and Malekzadeh, Reversed- phase liquid chromatographic method for the simultaneous determination of Antimalarial drugs Sulfadoxine, Pyrimethamine, Mefloquine and its major carboxylic metabolite in plasma, Journal of chromatography 571; 1991: 169-177.
10. Arnold PJ and Stetten OV, HPLC analysis of Mefloquine and its main metabolite by direct plasma injection with pre-column enrichment and column switching techniques, Journal of chromatography, 353: 1986: 193-200.
11. Willard HH, Merritt Jr. LL, Dean, JA, Settle, and Jr FA, “Instrumental Methods of Analysis”,., CBS Publishers and Distributors, Delhi, 2001, 7th ed, 1.
12. Skoog DA, Holler FJ, Timothy A and Nieman, NW, In; Principle of Instrumental Analysis, Eastern Press, Banglore, 2004, 5th Edn.,1-4, 729,731,732
13. United State Pharmacopoeia 27, National Formulary 22. Rockville, M.D., The United State Pharmacopoeia Convention, Inc 2004.
Received on 05.09.2011 Modified on 20.09.2011
Accepted on 24.09.2011 © RJPT All right reserved
Research J. Pharm. and Tech. 4(10): Oct. 2011; Page 1563-1566